Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–66
TLS 216 Programmer Manual
CURSor
?
CURSOR?
might return :CURSOR:FUNCTION OFF;MODE INDEPENDENT;VBARS:BINARY1
"---------------0";BINARY2 "---------------0";UNITS SECONDS;
POSITION1 500.0E-6;POSITION2 4.50E-3;SELECT CURSOR1;:CURSOR:
HBARS:UNITS BASE;POSITION1 3.20E+0;POSITION2 -3.20E+0;SELECT
CURSOR1;:CURSOR:PAIRED:UNITS BASE;POSITION1 500.0E-6;POSIĆ
TION2 4.50E-3;SELECT CURSOR1 as the current cursor settings.
CURSor:FUNCtion
Selects and displays the cursor type. Cursors are attached to the selected
waveform. This command is equivalent to setting Function in the Cursor menu.
Cursor
SELect:CONTROl
CURSor:FUNCtion { HBArs | OFF | VBArs | PAIred }
CURSor:FUNCtion?
FUNCtion
CURSor :
PAIred
<Space>
?
VBArs
OFF
HBArs
HBArs specifies horizontal bar cursors that measure volts.
OFF removes the cursors from the display.
VBArs specifies vertical bar cursors that measure time.
PAIred specifies paired cursors that measure both time and volts.
NOTE. HBArs and PAIred settings are not available if the selected group has a
display mode of Timing or Bus.
Examples
Group
Related Commands
Syntax
Arguments