Command Descriptions
2–164
TLS 216 Programmer Manual
IMMed
:
:
MEASUrement
CLOCk
:
SOUrce
<Space>
?
CH
GROUP
REF
REF
<x>
<y>
<z>
<z> CH <x>
CH<x> specifies that channel <x> be used as the waveform for the clock source.
GROUP<y> specifies that the channel currently selected in group <y> be used as
the waveform for the clock source. The command GROUP<x>:SELECT, described
on page 2–133, sets the selected channel.
REF<z> specifies that the currently selected channel in reference group <z> be
used as the waveform for the clock source. The command REF<x>:SELECT,
described on page 2–212, sets the selected channel.
REF<z>CH<x> specifies that channel <x> in reference group <z> to be used as the
waveform for the clock source.
MEASUREMENT:IMMED:CLOCK:SOURCE CH16
MEASUREMENT:IMMED:CLOCK:SOURCE?
might return REF1CH1, indicating the first channel of the first reference group
is the clock source.
MEASUrement:IMMed:DELay? (Query Only)
Returns the Delay measurement parameters for the immediate measurement.
(The immediate measurement is taken without displaying the results on screen.)
The delay parameters only affect delay type immediate measurements.
Measurement
MEASUrement:IMMed:DELay?
Arguments
Examples
Group
Syntax