Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–282
TLS 216 Programmer Manual
TRIGger:MAIn:PATtern:LIMIT? (Query Only)
Returns the main time-qualified trigger pattern parameters.
Trigger
TRIGger:MAIn:PATtern:LIMIT?
TRIGger
:
MAIn
PATtern
?
:
PATtern LIMIT
:
TRIGGER:MAIN:PATTERN:LIMIT?
might return :TRIGGER:MAIN:PATTERN:LIMIT:WHEN MORETHAN;WIDTH
2.0E-9.
TRIGger:MAIn:PATtern:LIMIT:WHEn
Sets or queries the definition of when the Boolean requirements must cease to
meet the conditions of the defined pattern to trigger the logic scope.
Trigger
TRIGger:MAIn:PATtern:LIMIT:WHEn { LESSThan | MOREThan | TIMEOut }
TRIGger:MAIn:PATtern:LIMIT:WHEn?
TRIGger
:
MAIn
PATtern
:
:
LIMIT
WHEn
:
MOREThan
TIMEOut
<Space>
?
LESSThan
LESSThan instructs the logic scope to trigger if the trigger inputs cease to meet
the Boolean requirements before the specified time.
MOREThan instructs the logic scope to trigger if the trigger inputs cease to meet
the Boolean requirements after the specified time.
TIMEOut instructs the logic scope to trigger following the specified time lapse, if
the trigger inputs still meet the Boolean requirements.
Group
Syntax
Examples
Group
Syntax
Arguments