Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–150
TLS 216 Programmer Manual
HORIZONTAL:MAIN?
might return :HORIZONTAL:MAIN:SCALE 1.0E-6.
HORizontal:MAIn:SCAle
Sets the time per division for the main time base. This command is equivalent to
setting Main Scale in the Horizontal Scale side menu.
Horizontal
HORizontal:DELay:SCAle, HORizontal:DELay:SECdiv, HORizon-
tal:MAIn:SECdiv
HORizontal:MAIn:SCAle <NR3>
HORizontal:MAIn:SCAle?
:
SCAle
HORizontal
:
MAIn
?
<Space> <NR3>
<NR3> is the time per division. The range is 5 s to 500 ps in a 1-2.5-5 sequence.
Values not set to the specified sequence will be adjusted to the closest valid
value.
HORIZONTAL:MAIN:SCALE 2.5E-6
sets the main scale to 2.5 ms per division.
HORIZONTAL:MAIN:SCALE?
might return 5E-9, which indicates the main time base is set to 5 nanosec-
onds.
HORizontal:MAIn:SECdiv
Sets the time per division for the main time base. This command is identical to
the HORizontal:MAIn:SCAle command. It is provided to maintain program
compatibility with some older models of Tektronix logic scopes.
Horizontal
Examples
Group
Related Commands
Syntax
Arguments
Examples
Group