Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
TLS 216 Programmer Manual
2–193
METHod
MEASUrement :
REFLevel
:
<Space>
?
ABSolute
PERCent
ABSolute specifies that the reference levels are set explicitly using the
MEASUrement:REFLevel:ABSolute commands. This method is useful when
precise values are required. For instance, when designing to published interface
specifications such as RS-232-C.
PERCent specifies that the reference levels are calculated as a percent relative to
HIGH and LOW. The percentages are defined using the MEASUrement:REFLe-
vel:PERCent commands.
MEASUREMENT:REFLEVEL:METHOD ABSolute
specifies that explicit user-defined values are used for the reference levels.
MEASUREMENT:REFLEVEL:METHOD?
returns either ABSolute or PERCENT, indicating the reference levels currently
in use.
MEASUrement:REFLevel:PERCent? (Query Only)
Returns the percent settings reference level.
Measurement
MEASUrement:REFLevel:PERCent?
MEASUrement
:
REFLevel
PERCent
:
?
MEASUREMENT:REFLEVEL:PERCENT?
might return a string similar to: :MEASUREMENT:REFLEVEL:PERCENT:HIGH
90.0E+0;LOW 10.0E+0;MID 50.0E+0;MID2 50.0E+0.