Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–238
TLS 216 Programmer Manual
TRIGGER:DELAY:PATTERN:INPUT 65535, 65535
sets all 16 channels to trigger on High input values and sets the Auxiliary
Input to “Don’t Care”.
TRIGGER:DELAY:PATTERN:INPUT 1, 1
sets channel 1 High and all other inputs to to “Don’t Care”.
TRIGGER:DELAY:PATTERN:INPUT?
might return 3,3, indicating channels 1 and 2 will trigger when their input
values are High. The other channels and the Auxiliary bits are set to “Don’t
Care”.
TRIGger:DELay:PATtern:LIMIT? (Query Only)
Returns the time-qualified pattern parameters for the delayed pattern trigger.
Trigger
TRIGger:DELay:PATtern:LIMIT?
TRIGger
:
DELay
PATtern
?
:
LIMIT
:
TRIGGER:DELAY:PATTERN:LIMIT?
might return :TRIGGER:DELAY:PATTERN:LIMIT:WHEN MORETHAN;WIDTH
2.0E-9.
TRIGger:DELay:PATtern:LIMIT:WHEn
Sets or queries the definition of “when” the Boolean requirements must cease to
meet the conditions of the defined pattern to trigger the logic scope.
Trigger
TRIGger:DELay:PATtern:LIMIT:WHEn { LESSThan | MOREThan
| TIMEOut }
TRIGger:DELay:PATtern:LIMIT:WHEn?
Examples
Group
Syntax
Examples
Group
Syntax