Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–96
TLS 216 Programmer Manual
NOTE. The DIAg:STATE EXECute command may take 30 seconds or longer to
respond. This command performs a warm boot and does not return control to the
instrument controller until diagnostics are complete.
Calibration and Diagnostic
DIAg:STATE EXECute
:
STATE
DIAg
<Space>
EXECute
EXECute runs the diagnostic test sequences specified by the DIAg:SELect
commands. When complete, the logic scope will return to its pretest state. If the
PON event was enabled before running the tests, a Service Request will be
generated. When the Service Request has been received, the pass/fail status of
the tests can be returned by executing the DIAg:RESUlt:FLAg? query.
The DIAg:STATE EXECute command clears the following:
H Event Queue
H Input Queue
H Status Registers (SESR and SBR)
To enable a power-on event to generate a Service Request, send the following
commands before running diagnostics:
H DESE 128
H *ESE 128
H *SRE 32
H *PSC 0
DIAg:STATE EXECUTE
executes all selected diagnostic tests.
DISplay? (Query Only)
Returns the current display settings.
Display
Group
Syntax
Arguments
Examples
Group