Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
TLS 216 Programmer Manual
2–183
GROUP<y> specifies that the channel currently selected in group <y> is the
waveform source for the measurement. The command GROUP<x>:SELECT,
described on page 2–133, sets the selected channel.
REF<z> specifies that the currently selected channel in reference group <z> is the
waveform source for the measurement. The command REF<x>:SELECT, described
on page 2–212, sets the selected channel.
REF<z>CH<x> specifies that channel <x> in reference group <z> is the waveform
source for the measurement.
MEASUREMENT:MEAS4:SOURCE2 CH1
sets channel 1 as the delay “to” source when making delay measurements.
MEASUREMENT:MEAS2:SOURCE2?
might return REF1CH3, indicating channel 3 of reference #1 is the current
source waveform for this measurement.
MEASUrement:MEAS<x>:STATE
Sets or queries the state (on or off) of MEAS<x>. MEAS<x> is one of the four
measurements available (named Measurement 1 through Measurement 4) where
<x> specifies 1 through 4. The source for the measurement can be selected using
the MEASUrement:MEAS<x>:SOURCE1 command.
Measurement
MEASUrement:MEAS<x>:STATE { OFF | ON | <NR1> }
MEASUrement:MEAS<x>:STATE?
MEAS
::MEASUrement
STATE
<x>
<Space>
OFF
?
ON
<NR1>
OFF or <NR1> = 0 turns the selected measurement off.
ON or <NR1> 0 turns measurements on.
MEASUREMENT:MEAS1:STATE ON
turns measurement defined as MEAS1 on.
Examples
Group
Syntax
Arguments
Examples