Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–176
TLS 216 Programmer Manual
MEASUrement:MEAS<x>:CLOCk:EDGe
Sets the clocking-edge slope of the clock waveform for MEAS<x>. MEAS<x>
is one of the four measurements available (named Measurement 1 through
Measurement 4), where <x> specifies 1 through 4. The clock waveform only
affects setup or hold measurement types. Use the MEASURE-
MENT:MEAS<x>:CLOCK:SOURCE command to select the clock source. This
command is equivalent to setting the clock edge in the Clock Edges and
Direction side menu.
Measurement
MEASUrement:MEAS<x>:CLOCk:EDGe { RISe | FALL }
MEASUrement:MEAS<x>:CLOCk:EDGe?
MEAS
:
:
MEASUrement
CLOCk
:
EDGe
<Space>
?
RISe
FALL
<x>
RISe specifies the rising edge.
FALL specifies the falling edge.
MEASUREMENT:MEAS<x>:CLOCK:EDGE RISE
specifies the rising clocking-edge slope of the clock waveform for
MEAS<x>.
MEASUREMENT:MEAS<x>:CLOCK:EDGE?
might return RISE.
MEASUrement:MEAS<x>:CLOCk:SOUrce
Sets or queries the source of the clock waveform for MEAS<x>. MEAS<x> is
one of the four measurements available (named Measurement 1 through
Measurement 4), where <x> specifies 1 through 4. The clock source only affects
setup or hold measurement types. This command is equivalent to setting the
clock source in the Source Setup side menu.
Group
Syntax
Arguments
Examples