Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
TLS 216 Programmer Manual
2–297
TRIGGER:MAIN:SEQUENCE:END:AUXILIARY HIGH
specifies that the AUX TRIGGER INPUT must be High, relative to the
threshold level, as part of the end pattern definition for the main sequence
trigger.
TRIGGER:MAIN:SEQUENCE:END:AUXILIARY?
might return 1, indicating that the end pattern definition for the main
sequence trigger is High.
TRIGger:MAIn:SEQuence:END:CH<x>
Sets or queries the logic level expected at the channel input defined by Ch<x> as
part of the end pattern definition for the main sequence trigger.
Trigger
TRIGger:MAIn:SEQuence:END:CH<x> { HIGH | LOW|X} NOTE: defines
CH<x> pattern bit to High, Low or Don’t Care
TRIGger
:
MAIn
SEQuence
:
:
END
<x>CH
:
LOW
X
<Space>
HIGH
HIGH specifies that the selected channel must be High relative to the threshold
level.
LOW specifies that the selected channel must be Low relative to the threshold
level.
X specifies that the selected channel can be either High or Low relative to the
threshold level.
TRIGGER:MAIN:PATTERN:CH14 HIGH
specifies that channel 14 must be High to meet the Boolean requirements of
the defined pattern for the main trigger.
TRIGGER:MAIN:PATTERN:CH11 HIGH?
might return LOW, indicating channel 11 must be Low to meet the Boolean
requirements of the defined pattern for the main trigger.
Examples
Group
Syntax
Arguments
Examples