Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–90
TLS 216 Programmer Manual
*RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp, TEKSecure
DELEte:SETUp { <NR1> | ALL }
<NR1>
DELEte :
SETUp
<Space>
ALL
<NR1> is a value from 1 to 10 specifying a setup storage location. Using an
out-of-range value causes an execution error.
ALL specifies all stored setups.
DELETE:SETUP ALL
removes all stored setups. The instrument initializes all ten storage locations
to the factory default setup.
DELEte:GROUP (No Query Form)
Deletes one or all of the stored reference waveform groups from memory;
memory allocated to the reference location then becomes available for realloca-
tion.
NOTE. The instrument does not clear waveform data from the reference location.
Save and Recall
RECAll:GROUP, SAVe:GROUP, TEKSecure
DELEte:GROUP { REF<x> | ALL }
DELEte :
<Space>
ALL
<x>
GROUP
REF
REF<x> (REF1, REF2, REF3 or REF) specifies one of the reference memory
locations.
ALL specifies all stored groups.
Related Commands
Syntax
Arguments
Examples
Group
Related Commands
Syntax
Arguments