Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–316
TLS 216 Programmer Manual
PATtern specifies that triggers occur when a specified pattern at the input
channels meet specified Boolean and time qualifications.
STATE specifies that a trigger occurs when specified patterns at the input
channels meet specified Boolean conditions at the time the clock input changes
state.
SEQuence specifies that a trigger occurs when a sequence of two patterns exist at
the input channels meeting specified Boolean and time qualifications.
TRIGGER:MAIN:TYPE EDGE
specifies EDGE (the normal trigger condition) as the main trigger type.
TRIGGER:MAIN:TYPE?
might return PULSE indicating that the main trigger type is a pulse trigger.
TRIGger:STATE? (Query Only)
Returns the current state of the triggering system.
Trigger
TRIGger:STATE?
STATE
TRIGger :
?
ARMed indicates that the instrument is acquiring pretrigger information. All
triggers are ignored when TRIGger:STATE is ARMING.
PARTial indicates that the main trigger has occurred and the logic scope is
waiting for trigger(s) for the delay by events.
REAdy indicates that all pretrigger information has been acquired and the
instrument is ready to accept a trigger.
TRIGger indicates that the instrument recognized a trigger and is acquiring
posttrigger information.
SAVe indicates that the instrument is in save mode and is not acquiring data.
AUTO indicates that the instrument is in auto mode and acquires data even in the
absence of a trigger.
Examples
Group
Syntax
Returns