Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–154
TLS 216 Programmer Manual
:
SCAle
?
<Space> <NR3>
HORizontal
<NR3> is the time per division. The range is 5 s to 500 ps in a 1-2.5-5 sequence.
Values not set to the specified sequence will be adjusted to the closest valid
value.
HORIZONTAL:MAIN 2.5E-6
sets the main scale to 2.5 ms per division.
HORIZONTAL:MAIN?
might return 5E-9, which indicates the main time base is set to 5 nanosec-
onds.
HORizontal:SECdiv
Sets or queries the time per division for the main time base and is identical to the
HORizontal:MAIn:SCAle command. It is included here for compatibility with
other Tektronix instruments.
Horizontal
HORizontal:SECdiv <NR3>
HORizontal:SECdiv?
:
SECdiv
?
<Space> <NR3>
HORizontal
<NR3> is the time per division. The range is 5 s to 500 ps in a 1-2.5-5 sequence.
Values not set to the specified sequence will be adjusted to the closest valid
value.
HORIZONTAL:SECDIV 2.5E-6
sets the main scale to 2.5 ms per division.
HORIZONTAL:SECDIV?
might return 5E-9, which indicates the main time base is set to 5 nanosec-
onds.
Arguments
Examples
Group
Syntax
Arguments
Examples