Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
TLS 216 Programmer Manual
2–265
as a 17-bit binary number. Bits 1 through 16 are used to set the “Don’t Care”
status of channels 1 through 16. Bit 17 sets the “Don’t Care” of the Auxiliary
Input. A bit value of zero will set the Boolean value to “Don’t Care” for that
channel, regardless of the first parameter bit value. A bit value of 1 for a channel
indicates the high or low value set by the previous parameter is used.
TRIGGER:DELAY:SEQUENCE:START:INPUT 65535, 65535
sets all 16 channels to trigger on High input values and sets the Auxiliary
Input to “Don’t Care”.
TRIGGER:DELAY:SEQUENCE:START:INPUT 1, 1
sets channel 1 High and all other inputs to to “Don’t Care”.
TRIGGER:DELAY:SEQUENCE:START:INPUT?
might return 3,3 indicating channels 1 and 2 will trigger when their input
values are High. The other channels and the Auxiliary bits are set to “Don’t
Care”.
TRIGger:DELay:SEQuence:STARt:LOGIC
Sets or queries the combinatorial logic pattern to apply to the delay sequence
trigger pattern.
Trigger
TRIGger:DELay:SEQuence:STARt:LOGIC { AND | OR | NAND | NOR }
TRIGger
:
DELay
SEQuence
:
:
STARt
:
LOGIc
<Space>
?
AND
OR
NAND
NOR
AND applies AND logic to the input channels. AND is a logic (Boolean) function
in which the output is true when and only when all the inputs are true.
OR applies OR logic to the input channels. OR is a logic (Boolean) function in
which the output is true if any of the inputs are true. Otherwise the output is
false.
Examples
Group
Syntax
Arguments