Tektronix TLS 216 Hunting Equipment User Manual


 
Command Descriptions
2–304
TLS 216 Programmer Manual
TRIGger
:
MAIn
SEQuence
:
:
LOWLimit
RANge
:
?
<Space>
<NR3>
<NR3> is the lower limit of the range in seconds.
TRIGGER:MAIN:SEQUENCE:RANGE:LOWLIMIT 1.0
sets the main sequence trigger lower limit range to 2 seconds.
TRIGGER:MAIN:SEQUENCE:RANGE:LOWLIMIT?
might return 100.0E-9, indicating the main sequence trigger lower limit
range is set to 100 ns.
TRIGger:MAIn:SEQuence:RANge:WHEn
Sets or queries when the channel inputs must cease to meet the Boolean
requirements of the pattern definition for the main sequence trigger.
Trigger
TRIGger:MAIn:SEQuence:RANge:WHEn { INRange | OUTRange }
TRIGger:MAIn:SEQuence:RANge:WHEn?
TRIGger
:
MAIn
SEQuence
:
:
RANge
WHEn
:
<Space>
?
OUTRange
INRange
INRange specifies that the trigger inputs must meet the Boolean requirements of
the pattern definition for the main sequence trigger for a period of time ending
within the user specified time range.
OUTRange specifies that the trigger inputs must meet the Boolean requirements
of the pattern definition for the main sequence trigger for a period of time ending
either before or after the user specified time range.
Arguments
Examples
Group
Syntax
Arguments