EVQty? (Query Only)
Returns the number of event codes in the Event Queue. Using this command is
useful when using the ALLEv? query since it lets you know exactly how many
events will be returned.
Status and Error
ALLEv?, EVENT?, EVMsg?
EVQty?
?
EVQty
<NR1>
EVQTY?
might return 3 as the number of event codes in the Event Queue.
FACtory (No Query Form)
Resets the logic scope to its factory default settings. This command is equivalent
to selecting Recall Factory Setup in the Waveform Save/Recall menu.
Miscellaneous
*PSC, *RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
FACtory
FACtory
Setting the logic scope to factory default:
H clears the Event Status Enable Register.
H clears the Service Request Enable Register.
H sets the Device Event Status Enable Register to 255.