GROUP<x>:VLOw?
VLOw
GROUP <x> :
?
<Space> <NR3>
<NR3> is the lower threshold from 12.00E–0 to 14.99E+0 volts.
GROUP1:VLOW 5.0E-1 sets the lower threshold voltage of group 1 to –0.5 volts.
GROUP2:VLOW?
might return 5.0E-1 to indicate the lower threshold voltage is set to –0.5
volts for group 2.
GROUP<x>:VOLts
Sets or queries the vertical gain of the specified group. This command is
identical to the GROUP<x>:SCAle command and is included for compatibility
purposes. Only GROUP<x>:SCAle is returned in response to a GROUP<x>?
query.
Vertical
GROUP<x>:SCAle
GROUP<x>:VOLts <NR3>
GROUP<x>:VOLts?
GROUP
<x> :
VOLts
?
<Space> <NR3>
<NR3> is the vertical gain from 50 mV to 2 V/div.
GROUP4:VOLTS 100E-03
sets the group 4 gain to 100 mV/div.
GROUP2:VOLTS?
might return 1.00E+0, indicating that the current V/div setting of group 2 is
1 V/div.