Figure 2Ć4:ăProbe Compensation Signal Displayed after AUTOSET
Passive probes require compensation to ensure maximum distortionĆfree
input to the TAS 455 and TAS 465 Oscilloscopes. Before taking any meaĆ
surements using a probe, first check the compensation of the probe and
adjust it to match the channel inputs. The signal source for this check is the
frontĆpanel PROBE COMP signal (Figure 2Ć5).
1. Attach the probe to either the CH 1 or CH 2 input connectors along the
lower right of the front panel.
2. Attach the probe tip to the PROBE COMP connector.
Figure 2Ć5:ăConnections for Compensating a Probe