Adjustment Procedures
Adjustment Procedures
5Ć18
5. Adjust the cursors until the DVolts readout is 6 V.
6. Note the cursor display error (are the cursors more or less than six
divisions apart).
7. Slide the cabinet off the instrument and adjust R112 on the Display
Driver board to compensate for the display error noted in step 6.
For example, if you noted in step 6 that with the DVolts readout at 6 V,
the cursor display equaled 6.2 divisions. Compensate by adjusting the
cursor display to 5.8 divisions with R112.
8. Slide the cabinet on the instrument and repeat this procedure until the
cursor display matches the DVolts readout.
Adjustment Complete
The adjustment procedure is complete. Install the cabinet and rear cover.