Tektronix TAS 465 Hunting Equipment User Manual


 
Warranted Characteristics
Specifications
1Ć14
TableĂ1Ć15:ăWarranted Characteristics Ċ XY Operation
Name
Description
XY Accuracy ±4%.
X Bandwidth DC to at least 3 MHz.
Phase Difference Between X and Y
Amplifiers
±3_, DC to 150 kHz.
TableĂ1Ć16:ăWarranted Characteristics Ċ Power Requirements
Name
Description
Source Voltage and Frequency 90-132 VAC
RMS
continuous range for 48 through 440 Hz.
180-250 VAC
RMS
continuous range for 48 through 440 Hz.
Power Consumption v85 watts (150 VA).
TableĂ1Ć17:ăWarranted Characteristics Ċ Environmental and Safety
Name
Description
Third Party Certifications This product shall conform to and be certified where appropriate
according to the following requirements:
Certification
Underwriter's Laboratories (UL) Listed or Recognized
Component, Electrical and
Electronic Measuring and Testing
Equipment UL 1244.
Canadian Standards CSA
Association (CSA)
Category Certified Electronic
Development and Test
Equipment, CSA 22.2 no. 231.
SelfĆCertification International
ElectroĆTechnical Commission
(IEC)
SelfĆcertified by Tektronix to
comply with IEC 348
recommendation.
Temperature: Operating and
Nonoperating
Operating: -10_ Cto+55_C.
Nonoperating: -51_ Cto+71_C.
Humidity: Operating and Nonoperating To 95% relative humidity at or below +40_ C to 75% relative
humidity form +41_ Cto+55_C.
A more sever test than stated in MIL-T-28800E, para. 4.5.5.1.1.2.
Altitude: Operating and Nonoperating Operating: To 4572 m (15,000 ft.).
Nonoperating: To 12192 m (40,000 ft.).
Random Vibration: Nonoperating 2.46 g
RMS
, from 5 to 500 Hz, 10 minutes each axis.