Tektronix TAS 465 Hunting Equipment User Manual


 
Removal and Installation Procedures
Maintenance
6Ć14
Power Supply to
Display Board
Display Board to
Display Board to
Analog Board
Power Supply to
Display Board
CPU to Front Panel
Analog Board
CPU to Analog Board
Display Control Board
to Display Board
Display Control Board
to Front Panel
Power Supply to CPU
Figure 6Ć4:ăCables and Cable Routing
The following procedures describe how to remove and install the line cord
and line fuse located on the back side of the oscilloscope.
WARNING
Unplug the line cord from the line voltage power source before
continuing. Failure to do so can cause injury or death.
Prepare the oscilloscope by positioning the bottom side down on the work
surface and facing the back side toward you. Refer to Figure 6Ć5.
Line Cord
1. Unplug the line cord from the power source.
2. Unplug the other end of the line cord from the back of the oscilloscope.
3. Grasp the end of the line cord and retaining clamp and rotate them
90 degrees counterclockwise (Figure 6Ć5).
4. Pull the line cord and clamp away from the back of the oscilloscope.
Line Cord and Line
Fuse