Performance Tests
Performance Verification
4Ć14
TableĂ4Ć1:ăRequired Test Equipment
Item Number and
Description
Minimum Requirements Example Purpose
1 Attenuator, 5X
1
Ratio: 5X; Impedance 50 W;
Connectors: female BNC input,
male BNC output
Tektronix part number
011-0060-02
Signal attenuation
2 Attenuator, 10X Ratio: 10X; Impedance 50 W;
Connectors: female BNC input,
male BNC output
Tektronix part number
011-0059-02
Signal attenuation
3 Termination,
50 W (two reĆ
quired)
Impedance 50 W; Connectors:
female BNC input, male BNC
output
Tektronix part number
011-0049-01
Signal
interconnection
4 Termination,
75 W
Impedance 75 W; Connectors:
female BNC input, male BNC
output
Tektronix part number
011-0102-01
Signal
interconnection,
video
5 Cable, Precision
Coaxial
50 W, 36 in, maleĆtoĆmale BNC
connectors
Tektronix part number
012-0482-00
Signal
interconnection
6 Cable, Coaxial 75 W, 36 in, maleĆtoĆmale BNC
connectors
Tektronix part number
012-1338-00
Signal
interconnection,
video
7 Coupler, DualĆ
Input
FemaleĆBNCĆtoĆdualĆmaleĆBNC Tektronix part number
067Ć0525Ć02
Signal
interconnection
8 Generator, LevĆ
eled Sine Wave
200 kHz to 250 MHz; Variable
amplitude from 5ĂmV to 4 V
pĆp
into 50 W
Wavetek 9100 Universal
Calibration System with
Oscilloscope Calibration
Module (Option 250)
Trigger and
bandwidth checks
9 Generator, Time
Mark
Variable marker frequency from
10 ms to 10 ns; accuracy within
25 ppm
M
o
d
u
l
e
(O
p
ti
on
250)
Timing accuracy
checks
10 Generator, Pulse High Amplitude pulse with variable
amplitude of 60 V to 100 V
Gain accuracy
checks
11 Generator, DC
Calibration
DC voltage levels from 100 mV to
10 V
Gain accuracy
checks
12 Probe, 10X, inĆ
cluded with this
instrument
Standard accessory probe TEKTRONIX P6109B Signal
interconnection
13 Digital
Multimeter
DC voltage to 10ĂV TEKTRONIX DMĂ2510 Power Supply
Adjustment
14 Generator, Video
Signal
Provides NTSC compatible
outputs
TEKTRONIX TSG 100 Video trigger check
15 Adjustment Tool
1
0.1 inch hex on both ends GC Electronics #8606 Internal adjustments
1
This item is used only when performing the Adjustment Procedures.