Tektronix TAS 455 Hunting Equipment User Manual


 
Warranted Characteristics
Specifications
1Ć12
TableĂ1Ć12:ăWarranted Characteristics Ċ Triggering System
Name
Description
TAS 465 EdgeĆType Trigger
S iti it DC C l d
The limits are as follows:
S
ensitivity, D
CC
oupled
Trigger Source
Sensitivity
Internal, Any Channel 0.30 division from DC to
25ĂMHz, increasing to 1 div at
150 MHz.
External X1 35 mV from DC to 25 MHz, inĆ
creasing to 100 mV at
150 MHz.
External X10 350 mV from DC to 25 MHz,
increasing to1Vat150 MHz.
TAS 455 EdgeĆType Trigger
Sensitivity DC Coupled
The limits are as follows:
Sensiti
v
it
y,
DC Co
u
pled
Trigger Source Sensitivity
Internal, Any Channel 0.30 division from DC to
25ĂMHz, increasing to 1 div at
100 MHz.
External X1 35 mV from DC to 25 MHz, inĆ
creasing to 100 mV at
100 MHz.
External X10 350 mV from DC to 25 MHz,
increasing to1Vat100 MHz.
Trigger Level or Threshold Accuracy The limits are as follows for signals having rise and fall
times w20 ns:
Coupling Source Accuracy
DC All except
Line and
External
±(5% of reading + 0.4 division
+ 1 mV)
External (X1) ±(5% of reading + 40 mV)
External (X10) ±(5% of reading + 400 mV)
Noise Reject All except Line
and External
±(5% of reading + 1.1 division
+ 1 mV)
External (X1) ±(5% of reading + 110 mV)
External (X10) ±(5% of reading + 1.1 V)
HF Reject All except Line
and External
±(5% of reading + 0.35 division
+ 1 mV)
External (X1) ±(5% of reading + 35 mV)
External (X10) ±(5% of reading + 350 mV)