Tektronix TAS 455 Hunting Equipment User Manual


 
Brief Performance Checks
Performance Verification
4Ć10
Gain and Position
Adjustment
Toggle Between
Gain and Position
Adjustment
Dots Eight Divisions
Apart, Centered
Horizontally
Figure 4Ć3:ăHorizontal Adjustments (One)
5. Select Done when you have the appropriate display.
6. Again, use a combination of the TOGGLE button and the General PurĆ
pose Knob to position the two dots eight divisions apart, centered
horizontally (see Figure 4Ć3).
7. Select Done when you have the appropriate display.
8. One dot is now displayed.
9. Use a combination of the TOGGLE button and the General Purpose
Knob to horizontally center the single dot at the center of the screen
(see Figure 4Ć4).
Continue to press the TOGGLE button and adjusting the dot to center
screen with the General Purpose Knob until the dot stays horizontally
centered while pressing the TOGGLE button.
Position
Adjustment
Toggle Dot Position
Center Dot
Horizontally
Figure 4Ć4:ăHorizontal Adjustments (Two)