Welch Allyn Medical Diagnostic Equipment CS Heart Rate Monitor User Manual


 
86 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor
Replacing the side panels
The Propaq CS monitor side panels are complete assemblies.
The side panels are sealed to the Rear Chassis with an adhesive gasket. A new gasket is
provided with the new side panel kit.
Removing the right side panel
1. Remove the battery pack (“Removing the single battery pack” on page 66).
2. Open the monitor (“Opening the monitor” on page 68).
3. Remove the Recharger Board (“Replacing the Recharger Board” on page 83).
4. Cut and remove the tie wrap securing the bead connected to the side panel.
5. Remove the two right side panel clamps by inserting the end of the flat blade
screwdriver under the clamp and gently rotating the blade to loosen the clamp. Then
slide the clamp out.
6. Press the side panel out of the Rear Chassis. There is an adhesive gasket around the
side panel which requires considerable force to separate. Do not pry against
components.
0696-42
INSERT SMALL SCREWDRIVER
BETWEEN PLASTIC AND CLAMP
TO RELIEVE TENSION AGAINST
PLASTIC