70 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor
P1 removal
6. Spread the casing open to an angle of about 45-60 degrees.
7. Detach the following cables from the Front Chassis (shown below):
• Cable to P5 on the Analog Board, from left side panel ECG connector on the Rear
Chassis.
• Cable to P6 on the Analog Board, from left side panel IPB connector on the Rear
Chassis.
• Cable to P7 on the Analog Board, from left side panel temperature jack on the
Rear Chassis.
• Cable to P8 on the lower (Digital) Board, from the right side panel on the Rear
Chassis.
• Cable to P9 on the Digital Board, from the right side panel on the Rear Chassis.
P6
P7
P5
P11
P10
P9
P7
P5
P6
P8
NOTE: You do not need to remove
the handle for this procedure. The
handle is not shown here so the
cable connections are more visible