Welch Allyn Medical Diagnostic Equipment CS Heart Rate Monitor User Manual


 
Service Manual Repair procedures 71
8. Detach the ribbon cable from the Rear Chassis (see below):
Ribbon cable to P10 on the Recharger Board, from the Digital Board.
cable p10, p11
9. Separate the Interface Chassis from the Rear Chassis (see below).
Note
The two halves are sealed by rubber gaskets that are recessed around the
perimeter of the Rear Chassis. If the gaskets become loose, replace them in the
perimeter before reassembling the monitor.
P10