Welch Allyn Medical Diagnostic Equipment CS Heart Rate Monitor User Manual


 
A
135
Manufacturable test equipment
The following device(s) can be manufactured for use in the functional verification and
calibration procedures in this manual. Parts are readily available from electronic
component suppliers.
CONNECTOR: 5-PIN, 0.156" CENTER
MFR: METHODE; PN 3300-405-220
PROTOCOL PN: 610-0036-00
(REQUIRES INSERTION TOOL
METHODE PN JT-11-156-ST)
1 2 3 4 5
USE RESISTOR NETWORKS
TO RIGHT AND CREATE TWO (2)
SEPARATE TEST LOADS:
4.9 OHM AND 4.65 OHM LOADS
RESISTIVE TEST LOADS
(USED TO ADJUST
RECHARGER POWER SUPPLIES)
5 Ω
20W
5 Ω
20W
.2 Ω
2W
.2 Ω
2W
68 Ω
5W
20 Ω
5W
39 Ω
5W
5 Ω
5W
4.9 Ω
4.65 Ω
Welch Allyn PN: 610-0036-00