Welch Allyn Medical Diagnostic Equipment CS Heart Rate Monitor User Manual


 
Service Manual Repair procedures 97
Opening the SpO
2
Module
1. Separate the SpO
2
module from the monitor according to the procedure “Removing
the dual battery in expansion module” on page 66.
2. Two tabs secure the SpO
2
module rear panel to the SpO
2
module chassis. Using a
small screwdriver or your finger, press on each of the tabs inside the SpO
2
module
chassis while applying pressure to separate the rear panel from the chassis. (When
separating the rear panel from the chassis, make sure the side panels remain seated
in the rear panel.)
spo2 chassis disassy
3. After the tabs allow the rear panel to be removed, separate the rear panel from the
chassis.
The rear panel is sealed to the chassis with rubber O-ring gaskets. These gaskets can
be reused when re-assembling the SpO
2
module.
When closing the SpO
2
module, confirm that no O-ring gasket slips out of its channel and
becomes pinched between the rear panel and the Chassis.
SpO2 Module
Rear Panel
Release
Tabs
Continuous
O-Ring Gasket
0696-53
Rear
Panel
O-Ring
Cord
Release
Tab (2)
SpO2
Board
SpO2
Connector
Panel
Chassis
Speaker
SCP Board