6 Overview Welch Allyn Propaq CS Vital Signs Monitor
Warranty service
DO NOT ATTEMPT TO REPAIR the monitor yourself during the warranty period. For
service and repair, contact Welch Allyn. See below.
Technical support services
Welch Allyn offers a wide range of technical support services including:
24-hour telephone support
loaner equipment
service contracts
field replaceable units (FRUs)
factory repair
For any of these services, contact Welch Allyn Technical Support:
Worldwide Internet Access
World Wide Web (WWW) URL: http://www.monitoring.welchallyn.com/service
World Wide E-mail: solutions@monitoring.welchallyn.com
Recommended service intervals
At the intervals listed below, check the Propaq CS monitor for normal operation.
USA 1-800-535-6663 France (+33) 1-60-09-33-66 Australia (+61) 2-9638-3000
Latin America (+1) 305-669-9591 Germany (+49) 7477-927-173 Singapore (+65) 6291-0882
European Call
Center
(+353) 469-067-790 Canada 1-800-561-8797 Japan (+81) 3-5212-7391
United Kingdom 0-207-365-6780 South
Africa
(+27) 11-777-7509 China (+86) 21-6327-9631
Interval/Condition Perform Located in this
manual
Every 6 months to 2 years (according to
hospital protocols).
Complete risk (leakage) current Safety Check
followed by a Functional Verification.
“Functional
verification”
If battery does not retain a charge. Check battery pack capacity. “Troubleshooting”
Monitor is dropped or suspected of
damage or rough handling.
Complete Safety Check followed by Functional
Verification.
“Functional
verification”
Suspected malfunction with all or part of
monitoring parameters.
Functional Verification of suspected parameter(s). “Functional
verification”
Monitor does not pass Functional
Verification.
Calibration of monitoring parameter circuits, or
repair, followed by Safety Check and Functional
Verification.
“Functional
verification” and
“Calibration”
WARNING If the monitor is opened for repair or calibration, a dielectric strength
test must be completed to ensure the integrity of the patient isolation barrier. The
dielectric strength test procedure is provided in “Functional verification”.