66 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor
Removing the single battery pack
The following steps describe how to remove the battery pack from the monitor:
1. Using a flat blade screwdriver, unscrew the four screws securing the battery pack
cover as shown below.
2. Disconnect the monitor cable from the battery pack cable.
3. Remove the battery pack from the monitor.
4. Store the battery in a safe place while disassembling the monitor.
Removing the dual battery in expansion module
The Expansion Module uses a dual battery pack. The following instructions describe how
to remove the dual battery.
1. Using a Phillips screwdriver, remove the four screws securing the Expansion Module
to the monitor (see below).
Monitor cable
Battery pack
Battery pack cable
Battery pack cover
Note
Four screws secure the Expansion Module to the monitor. Four additional screws
secure the rear panel to the Expansion Module Chassis; do not remove these
additional screws.