Fluke PM3370B Hunting Equipment User Manual


 
4 - 50 HOW TO USE THE INSTRUMENT
MEASURE MEAS1-PKPK
- Press the ANALOG key to select the digital mode.
- Press the MEASURE key.
The displayed menu gives access to the two measurements MEAS 1 and
MEAS 2. Each measurement can be independently turned on and off. In this
menu, you can select the measurement in MEAS 1 and MEAS 2 function.
Observe that the measurements are linked to the active channel. The two default
measurements are ’pkpk’ and ’freq’.
- Press the second softkey to turn MEAS1 to ’on’.
The screen automatically displays the result of the peak-to-peak measurement:
’ch1: pkpk= ....mV’.
- Press the softkey next to ’MEAS1’.
The menu displays the ’SELECT MEAS 1" submenu. You can choose from three
main measurements: volt, time, or delay. The lower part of the menu optimizes to
the selected main measurement.
The actual type of measurement is selected by using the ’TRACK’ control. The
following voltage measurements are available : dc, root- mean square, minimum
peak, maximum peak, peak-to-peak, low level, high level.
The following time measurements are available : frequency, period, pulse width,
rise time, fall time, duty cycle.
The ’delay’ measurements include the time delay between leading or trailing
edges between any two channels or traces. (Channel or trace
must
be on the
display)
MEASURE MEAS1-RMS
- Select ’volt’ measurements.
- Turn the TRACK control to select ’rms’.
Observe that the screen displays the result of the rms measurement: ’ch1 rms = ....V’.