Fluke PM3370B Hunting Equipment User Manual


 
6 - 22 THE CPL PROTOCOL
PROGRAM WAVEFORM PW
Purpose:
To send a waveform to the oscilloscope. This function is referred to as to
P
rogram
a complete
W
aveform in the oscilloscope.
A waveform is sent (programmed) in two command sequences. The first
sequence selects the waveform register number, programs the waveform
administration data, and programs the number of samples. The second sequence
programs the samples, including a checksum. Each sequence is synchronized by
an acknowledge response.
You can use the
PW
command to send the data received with the
QW
command
back to the oscilloscope.
Note: The number of data bits of the RS-232 interface must be 8 (not 7),
otherwise, an execution error will occur.
Command 1:
Response 1:
Command 2:
Response 2:
"PW" pds wave_nr pds admin CRpds count
acknowledge CR
sample checksum
acknowledge CR