Philips M3500B Heart Rate Monitor User Manual


 
Defibrillating
M3500B HeartStart XLT Defibrillator/Monitor 6-5
6
Figure 6-4 Charging Display
If desired, you may increase or decrease the selected energy level after
pressing the button. To do so, press repeatedly until the
desired energy level is displayed. The defibrillator charges to the selected
energy automatically. Wait until the current charge reaches the selected
energy level before proceeding.
3. Shock
Confirm that a shock is still indicated. Make sure no one is touching the
patient or anything connected to the patient. Call out “Clear.” Then press
to deliver a shock to the patient.
To disarm the defibrillator, press . If is not pressed in 30 sec-
onds, the defibrillator disarms automatically.
If additional shocks are indicated, repeat the defibrillation process
WARNIN G Defibrillation current can cause operator or bystander injury. Do not touch the
patient, or equipment connected to the patient, during defibrillation.