Philips M3500B Heart Rate Monitor User Manual


 
System Messages
12-4 Troubleshooting
6S2 &DEOH2II The SpO
2
cable is not connected to the
device.
Attach the SpO
2
cable to the HeartStart
XLT.
632
)DLOXUH A failure has occurred in the SPO
2
circuitry.
Remove the device from active use and
call for service.
6S2
/LJKW,QWHUI The level of ambient light is so high that
the sensor cannot obtain an SpO
2
reading
or the sensor or cable is damaged.
Cover the sensor with an opaque
material.
Check the sensor for damage; try
another sensor.
6S2
1RQ3XOVDWLOH The patient’s pulse is absent or too weak
to be detected.
Check that the sensor is applied prop-
erly.
Make sure the sensor site has a pulse.
Relocate the sensor to a site with
improved circulation.
Try another sensor.
6S2 /RZ6LJQDO SpO
2
signal is too low to give an accurate
reading.
Check the sensor is applied properly.
Try another sensor type.
6S2
1RLV\6LJQDO Excessive patient movement, electrical
interference, or optical interference is
present.
Minimize patient movement or apply
the sensor to a site with less move-
ment.
Secure the sensor cable loosely to the
patient.
Reduce sources of electrical or optical
interference.
6S2
6HQVRU)DLO The SpO
2
transducer is broken.
Try another sensor.
Table 12-1 System Messages (Continued)
Message Description Corrective Action