System Messages
M3500B HeartStart XLT Defibrillator/Monitor 12-5
12
Momentary Messages
Momentary messages are temporary and only appear on the display for a few
seconds. Each message is accompanied by a three second beep to alert you.
Table 12-2 lists momentary messages.
Table 12-2 Momentary Messages
Message Possible Cause Corrective Action
$WWDFK3DGV The multifunction defib electrode pads
are not making proper contact with the
patient.
Check the pads are applied to the
patient, as directed on the pads’ pack-
age.
Replace the pads if the prompt contin-
ues.
$WWDFK/HDGV An attempt was made to begin pacing in
Demand Mode without ECG leads
attached to the patient.
Attach leads to patient.
'HILE'LVDUPHG
The pads connection is compromised.
The mode is changed from Manual to
AED while the defibrillator is
charged.
is not pressed within 30 sec-
onds of the defibrillator being
charged.
is pressed.
Check the pads are applied to the
patient properly.
If a shock is indicated, deliver the
shock before changing modes.
To deliver a shock, press
within 30 seconds of the defibrillator
being charged.
1R6KRFN'HOLYHUHG Patient impedance is too high.
Make sure the pads are applied prop-
erly.
Replace the pads, if necessary
&KHFN3ULQWHU Printer paper is absent or jammed; the
printer door is not closed properly.
Reload printer paper.
Make sure the door is closed properly.
6+2&.
',6$50
6+2&.