Applying the Sensor
M3500B HeartStart XLT Defibrillator/Monitor 5-5
5
Applying the Sensor
Follow the manufacturer’s directions for applying and using the sensor, mak-
ing sure to observe any warnings or cautions. For the best results:
Make sure the sensor is dry.
If the patient is moving, secure the sensor cable loosely to the patient.
Avoid excessive pressure at the sensor site; ensure that circulation is
not obstructed.
Keep power cables away from the sensor cable and connection.
Avoid placing the sensor in an environment with bright lights (if nec-
essary, cover the sensor with opaque material).
Avoid placing the sensor on any extremity with an arterial catheter,
blood pressure cuff, or intravascular venous infusion line.
WARNIN G Failure to apply the sensor properly may reduce the accuracy of the SpO
2
measur-
ment.
WARNIN G Inspect the sensor application site at least every two hours for changes in skin qual-
ity, correct optical alignment, and proper sensor application. If skin quality is com-
promised, change the sensor site. More frequent checking may be required due to an
individual patient’s condition.