Configuring the HeartStart XLT
10-12 Setting Up and Configuring the HeartStart XLT
NOTE * If (XURSHDQ3URWRFRO is set to 2II, the Pause Timer is used during the Pause state
and appears as a configurable parameter. If (XURSHDQ3URWRFROis set to 2Q, either
the 3RVW6KRFN&357LPHU or the ´16$µ7LPHU are used during the Pause state and
appear as configurable parameters in place of 3DXVH7LPHU.
NOTE If (XURSHDQ3URWRFRO is set to 2Q, the setting for the 6KRFN6HULHV7LPHU must be <
3RVW6KRFN&357LPHU <
16$7LPHU. Also, the 6KRFN6HULHV7LPHU can not be config-
ured to either 2II or .
16$7LPHU Defines the duration of the Pause time (in sec-
onds) when is pressed and the time
since the last shock is greater than the 6KRFN
6HULHV7LPHU setting - typically when 1R6KRFN
$GYLVHG
30, 60, 120,
180
If (XURSHDQ3URWRFRO is set to Off, the following configuration choice appears:
´3DXVHµ7LPHU Defines the duration of the pause time (in sec-
onds), when is pressed (when (XURSHDQ
3URWRFRO is set to Off).
30, 60, 120,
180
The following two configuration choices are always available regardless of the
(XURSHDQ3URWRFRO setting.
/HDG6HOHFW Turns on ECG monitoring. On, Off
6S2 Turns on SpO
2
monitoring. Only listed as a
configurable item if the SpO
2
option was pur-
chased.
On, Off
Table 10-2 AED Settings: (Continued)
Parameter Description
Setting
Choices
3$86(
3$86(