Specifications
13-4 Specifications & Safety
ECG Monitoring
Inputs: Single channel ECG may be viewed on display and printed. PADS
ECG is obtained through two multifunction defibrillation electrode pads.
Lead I, II, or III is obtained through the 3-lead ECG cable and separate moni-
toring electrodes. With a 5-lead cable, lead aVR, aVL, aVF, or V can also be
obtained.
Lead Fault: LEADS OFF message and dashed line appear on the monitor if a
lead electrode or wire becomes disconnected.
Pad Fault: PADS OFF message and dashed line appear on the monitor if a
pad becomes disconnected.
Heart Rate Display: Digital readout on display from 15 to 300 bpm, with an
accuracy of 10%.
Heart Rate Alarms: Configurable pairs of low and high heart rate alarm lim-
its: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.
Defibrillation Patient Cable Length: 7 ft (2.13 m).
ECG Cable Length: 10 ft (3.05 m).
Common Mode Rejection: Greater than 90 dB measured per AAMI stan-
dards for cardiac monitors (EC 13).
ECG Size: 2.5, 5, 10, 20, 40 mm/mV.
Frequency Response:
AC Line Filter: 60 Hz/50Hz.
Pads ECG for Display: Monitor (.15-40 Hz)/EMS (1-30 Hz).
Pads ECG for Printer: Monitor (.15-40 Hz)/EMS (1-30 Hz).
Leads ECG for Display: Monitor (.15-40 Hz)/EMS (1-30 Hz).
Leads ECG for Printer: Diag (.05-150 Hz)/EMS (1-30 Hz)/
Monitor (.15-40 Hz).
Patient Isolation (defibrillation proof):
ECG: Type CF
SpO
2
: Type CF
Defib: Type BF (multi-function defib electrode pads)
+
_