Configuring the HeartStart XLT
M3500B HeartStart XLT Defibrillator/Monitor 10-11
10
Table 10-2 AED Settings:
Parameter Description
Setting
Choices
$('6KRFN6HULHV Defines the maximum number of shocks to
deliver before prompting &KHFN3DWLHQW&KHFN
3XOVH,I1HHGHG%HJLQ&35.
2, 3, 4
6KRFN6HULHV7LPHU Defines the number of seconds that must pass
before the next shock is considered the first
shock of a new shock series, rather than the
next shock of the current shock series.
30, 60, 90,
120, 150, 180,
210, Off
'HYLFH,QLWLDWHG
$QDO\VLV
Initiates ECG analysis when the HeartStart
XLT is turned on in AED Mode.
On, Off
$XWRPDWLF
5H$QDO\VLV
Initiates ECG analysis in between shocks
within a shock series.
On, Off
5K\WKP0RQLWRULQJ Monitors the ECG for potentially shockable
rhythms when the HeartStart XLT is not ana-
lyzing, defibrillating, or paused.
On, Off
´&KHFN3DWLHQWµ
7LPHU
Defines how often (in seconds) the &KHFN
3DWLHQW prompt is repeated when Rhythm Mon-
itoring detects a potentially shockable rhythm.
30, 45, 60, 90,
Off
(XURSHDQ3URWRFRO Modifies Pause state prompts and replaces the
Pause Timer with either the Post Shock CPR
Timer or the "NSA" Timer, depending on the
event preceding the Pause state.
Off/On
If (XURSHDQSURWRFRO is set to On, the following two configuration choices appear:
3RVW6KRFN&35
7LPHU
Defines the duration of the Pause time (in sec-
onds) when is pressed and the time
since the last shock is less than or equal to the
6KRFN6HULHV7LPHU setting - typically at the com-
pletion of a shock series.
30, 60, 120,
180
3$86(