Philips M3500B Heart Rate Monitor User Manual


 
Defibrillating
3-10 Defibrillating in AED Mode
3. If shock advised, press .
Once charging is complete, the charge tone becomes continuous. Make
sure no one is touching the patient or anything connected to the patient.
Call out "Clear." Then press to deliver a shock to the patient.
Figure 3-8 Press SHOCK Display
WARNIN G Defibrillation current can cause operator or bystander injury. Do not touch the
patient, or equipment connected to the patient, during defibrillation.
The defibrillator automatically disarms if you do not press in 30
seconds.
6+2&.
6+2&.
Shocks: 0
00:00:14
Stand Clear!
Press SHOCK
150J
SHOCK
6+2&.