Service Manual Repair procedures 69
10. Re-insert the rubber gasket (part of the fastener kit) into the groove in the front case.
Begin inserting the gasket at the top of the rear case in the four-turn groove area.
Position the rubber gasket flat in the groove area of the front case.
11. Insert the ECG sensor cable into the J1 connector and lock the latching piece.
Installing the main board, subframe and radio card (2.4 GHz)
1. Install the main board, subframe and radio card as one assembly onto the front case.
Caution Do not to bend or tilt any of the neon bulbs. They must remain vertical.
Four-turn groove area
ECG sensor flex cable at J1
Neon bulbs