Welch Allyn Medical Diagnostic Equipment Model 406 Heart Rate Monitor User Manual


 
20 Functional verification Welch Allyn Micropaq Monitor
SpO
2
information
Network Monitor
Use the Network Monitor screen as a diagnostic aid for troubleshooting wireless
connections between the monitor and a network access point.
SpO
2
Information screen examples
TX PACKETS PKTS: Total number of packets sent
DIR: Total number of directed packets sent (2.4 GHz radio card) or
“N/A” (5 GHz card)
ERRORS: Total number of TX packet errors
(xxx%) The calculated percentage of errors during the last update
period
RX PACKETS PKTS: Total number of packets received
DIR: Total number of directed packets received
ERRORS: Total number of RX packet errors
(xxx%): The calculated percentage of errors during the last update
period
PSICP TX / RX Total number of PSICP packets sent and received in this PSICP
session. Welch Allyn internal use only.
Network Monitor screen