Welch Allyn Medical Diagnostic Equipment 204el Heart Rate Monitor User Manual


 
Temperature
Propaq Encore Reference Guide 67 Welch Allyn
Temperature Messages
The following messages can appear in an equipment alert window. A temperature caution
message will also appear in the temperature numeric window when one of these messages
appears (except PROBE NOT DETECTED).
PROBE NOT DETECTED. This message occurs when the Propaq Encore has
successfully measured temperature and a probe is then disconnected. Reconnect the
probe or acknowledge the equipment alert by pressing any menu button.
PROBE SHORT. Verify that the probe is properly inserted in the left side panel. If so,
replace probe.
CALIBRATION ERROR, TEMP DISABLED. This message appears when the Propaq
Encore has detected that it cannot accurately measure the temperature. The monitor
should be serviced.
Malfunction of the temperature probes may result in inaccurate readings. Confirm suspect
readings.