Welch Allyn Medical Diagnostic Equipment 204el Heart Rate Monitor User Manual


 
Monitor Setup
Propaq Encore Reference Guide 31 Welch Allyn
Selecting Waveforms for Display
To select waveforms for display, press SETUP, WAVE SEL. Use the NEXT and ON/OFF buttons to turn
on the desired waveforms in the wave select window:
Display Priorities
You can turn on more than three waveforms, but only the first three waveforms listed in the
wave select window that are monitored are displayed. The patient parameters being
monitored are listed in the order they will be displayed if all are turned on.
Because of the critical nature of the ECG waveform, you cannot turn off ECG. However, if
ECG is not monitored, another waveform will occupy its place.
The displayed waveforms are also the ones printed if a printer is attached.
Setting the Time and Date
To set the time and date, from the Main Menu press SETUP, MORE, MORE. The monitor displays
the Time/Day window:
Press
NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER to store the new
time and date.
WAVE SELECT
mmH
g
NEXT PREV MENUINSERVON/OFF
MCO
2 BR
S
p
O
2
PA
SETUP
ECG : ON RESP : OFF
ART : ON S
p
O2 : ON
PA : OFF NIBP : ON
CO
2 : OFF
35/ 18
( 85)
35
12
92
BATTERY: 9.2 VOLTS
mmH
g
NEXT PREV MENUDOWN ENTERUP
MCO
2 BR
S
p
O
2
PA
TIME/DAY
TIME DAY
H:MIN:S MO/DA/YR
07:45:32 06/12/97
35/ 18
( 85)
35
12
92