NIBP
Welch Allyn 58 Propaq Encore Reference Guide
If artifact is so severe while Smartcuf is enabled that it affects the accuracy of an NIBP
measurement, that measurement is marked with a special symbol on the display and on
printouts
There may be some situations where it is desirable to disable Smartcuf. This may include
situations with very extreme motion artifact, certain types of arrhythmias, or other situations
where it is not possible to obtain a good ECG signal. NIBP measurements can still be
performed when Smartcuf is disabled.
To disable Smartcuf, from the Main Menu press
NIBP to display the NIBP Menu and set
Smartcuf to OFF.
NIBP Connector and Cuffs
Use Propaqs with approved hoses and cuffs listed in the Welch Allyn Products and Accessories
booklet. The Propaq Encore uses a single-hose cuff. Cuffs that conform to AAMI or AHA
guidelines should be used. Select the proper size of cuff based on the limb circumference.
Neonate Pediatric Adult
Hoses Neonate/Infant Adult Adult
Cuffs (typical cuff
labeling)
Neonate #1 to #5
(disposable); newborn,
infant (reusable)
Newborn, infant, small
child, child, small adult
child, small adult, adult,
large adult, thigh
Recommended limb
circumference
up to 15 cm 7.7 to 25 cm greater than 15 cm
mmH
g
MCO
2 BR
S
p
O
2
35/ 18
( 85)
35
12
92
PRINT OXYCRGNXT TRND
NIBP TREND
TIME BRSYS/DIA-MEANHR/PR
BPM
Br/MHH:MM
10:01
9:46
9:31
9:20
98
SRCH
OFF
100
101
88
93
91
11
10
5
11
139/ 74 106
142/ 83 110
145/ 87 112
126/ 85 91
S
p
O2
% mmH
g
PA
mmHg
122
58
85
This symbol indicates the
NIBP reading was taken in
the presence of high
artifact while monitoring
ECG with Smartcuf
enabled. Artifact can affect
accuracy.
Trend Display
Printout
Numeric Display