Monitor Setup
Propaq Encore Reference Guide 29 Welch Allyn
Monitor Setup
Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu.
STATSCALE
Automatically readjusts all waveform scales.
ALARMS Allows access to the Alarms menu.
WAVE SEL Allows you to turn on and off desired waveforms or NIBP
numerics for display.
TRENDS
Allows access to the Trend settings and display.
MORE
Displays the next setup menu and the following status window:
NEXT
Selects the next setting in the status window.
CHANGE
Changes the currently selected display setting. (Pressing CHANGE at
PATIENT MODE allows you to choose between Adult, Pediatric, and
Neonatal in a Patient Mode window.)
PRINTER Allows access to the Printer Menu.
MORE
Allows access to the Time/Day window.
CURRENT SOURCE When the selected HR/PR source is no longer available, the current
source is the active source with highest priority. The RR/BR source
cannot be manually selected. It will always be CO
2
if CO
2
is active.
Otherwise, it will be ECG/RESP.
SELECTED SOURCE The user-selected HR/PR source is displayed along with the HR/
PR source currently being used by the monitor.
SWEEP (mm/s) The selectable sweep speeds for HR/PR are 12.5, 25, and 50 mm/
sec. The sweep speeds for RR/BR are 3.13, 6.25, and 12.5 mm/sec.
ALARM TONE Sets the Alarm Tone volume to HIGH, MEDIUM, or LOW.
STATSCALE MOREWAVE SEL TRENDSALARMS
mmH
g
NEXT MOREWAVE SEL PRINTERCHANGE
MCO
2 BR
S
p
O
2
PA
SETUP HR/PRHR/PR RR/BR
CURRENT SOURCE : ECG MCO2
SELECTED SOURCE : ART ---
SWEEP (mm/s) : 12.5 3.13
ALARM TONE : HIGH
HR/PR TONE : LOW
PATIENT MODE : ADULT
35/ 18
( 85)
35
12
92