Monitor Setup
Welch Allyn 34 Propaq Encore Reference Guide
From the Main Menu, press SETUP, MORE, CHANGE, SETUP. The Mode Setup window appears:
The asterisk (*) indicates which patient mode is currently selected for powerup.
Whenever you change the patient mode, the alarm limit settings automatically
change to the settings for that mode.
NEXT
Selects the next setting in the status window.
POWERUP*
Selects the highlighted patient mode (and its associated settings) as the
powerup mode. The selected powerup mode is marked by an asterisk (*).
(This does not change the current patient mode.)
USE NOW
Selects the highlighted patient mode (and its associated settings) as the
current patient mode. (This does not affect the powerup mode.)
SAVE
Use this button to reprogram the settings of Custom patient mode:
1.
Make sure the patient mode you want to reprogram (ADULT, PED or NEO) is
currently-used (to change patient modes, highlight the desired mode and
press USE NOW, YES).
2. Exit the Mode Setup window, then use other menus and buttons to set the
monitor settings and alarm limits as desired.
3. Re-enter the Mode Setup window, highlight the desired Custom mode, and
press SAVE, YES.