Welch Allyn Medical Diagnostic Equipment 204el Heart Rate Monitor User Manual


 
NIBP
Welch Allyn 142 Propaq Encore Reference Guide
NIBP
NIBP Specifications
Characteristic Specification
Method Oscillometric
Control Automatic and manual measurement control
Auto Intervals 1, 2, 3, 5, 10, 15, 30, and 60 minutes
Turbocuf Maximum measurements allowable in a 5-minute period
Displayed Pressures Systolic, Diastolic, and Mean plus on-screen manometer
Systolic Range Adult: 30 to 260 mmHg
Ped: 30 to 160 mmHg
Neonate: 25 to 120 mmHg
Diastolic Range Adult: 20 to 235 mmHg
Ped: 15 to 130 mmHg
Neonate: 10 to 105 mmHg
Mean Range Adult: 20 to 255 mmHg
Ped: 15 to 140 mmHg
Neonate: 10 to 110 mmHg
Static Manometer Accuracy ±3 mmHg
Minimum Inflation Pressure Adult: 100 mmHg
Ped: 80 mmHg
Neonate: 50 mmHg
Maximum Allowable Pressure Adult: 270 mmHg
Ped: 170 mmHg
Neonate: 132 mmHg
Default Inflation Pressure Adult: 160 mmHg
Ped: 120 mmHg
Neonate: 90 mmHg
Normal Overpressure Limit (results in
up to 2 retries)
Adult: 280 mmHg
Ped: 200 mmHg
Neonate: 141 mmHg
Single Fault Overpressure Limit Adult: 308 mmHg
Ped: 220 mmHg
Neonate: 154 mmHg
Leak Rate After a 1 minute settling period, leak rate is 4mm/Hg over a 3-minute
period at 270 mm/Hg.
Pulse Rate Range 30 to 220 beats per minute
Maximum Determination Time (with
retries)
Adult: 4.5 minutes
Ped: 4 minutes
Neonate: 3 minutes
Maximum Determination Time (no
retries)
Adult: 3 minutes
Ped: 2 minutes
Neonate: 1.5 minutes
Typical Determination Time without
Artifact
30 to 45 seconds