Welch Allyn Medical Diagnostic Equipment 204el Heart Rate Monitor User Manual


 
Defib Sync Message
Welch Allyn 136 Propaq Encore Reference Guide
7. Follow hospital procedures and LIFEPAK 6s instructions for cardioversion.
8. If subsequent cardioversion must be performed, repeat steps 5 through 7.
Defib Sync Message
The Defib Sync message can appear in an equipment alert window.
DEFIB FAULT/CHECK INTERFACE CABLE indicates a probable shorted cable.
Warning If the R-wave synchronization markers do not appear to be nearly
simultaneous with the R-wave peaks on the Propaq Encore display or are not
present, do not proceed with synchronized cardioversion.
You must press the LIFEPAK 6s SYNC button and check for appropriate
synchronization markers on the Propaq Encore before each cardioversion.
Welch Allyn cannot guarantee the delay from the sync marker to the
defibrillator discharge.
A fault in the cable between the Propaq Encore and the defibrillator, or
unplugging the cable, will prevent showing markers on the Propaq and can
prevent setting the defibrillator to synchronized mode.